Wiring Prosonic T
12 Endress+Hauser
3 Wiring
"
Caution!
Before connection please note the following:
• The power supply must be identical to the data on the nameplate.
• Switch off power supply before connecting up the instrument.
• Connect equipotential bonding to devices ground terminal before connecting up the
instrument → ä 14, "Potential matching".
#
Warning!
When you use the measuring system in hazardous areas, make sure to comply with national
standards and the specifications in the safety instructions (XA’s). Make sure you use the specified
cable gland.
3.1 Wiring
1. Unscrew housing cover (1).
2. Remove display (2) if fitted.
3. Insert cable (3) through gland (4).
"
Caution!
If possible, insert the cable from
above and let a draining loop in
order to avoid intrusion of
humidity.
4. Installation cable screen to the
grounding terminal (5) within the
terminal compartment.
5. Make connection according to
terminal assignment, → ä 13,
"Terminal assignment".
6. Tighten cable gland (4).
7. Insert display (2) if fitted.
8. Screw on housing cover (1).
9. Switch on power supply.
L00-FMU30KAx-04-00-00-xx-008
1
2
1
2
3
5
4