Prosonic T Wiring
Endress+Hauser 13
3.2 Terminal assignment
L00-FMU30xxx-04-00-00-en-015
3.3 Supply voltage
The following values are the voltages across the terminals directly at the instrument:
Current
consumption
Terminal voltage minimum
Terminal voltage
maximum
4 mA 14 V 35 V
20 mA 8 V 35 V
3
1
2
L-
L+
4...20 mA
plant
ground
power