EasyManua.ls Logo

Endress+Hauser Prothermo NMT81 - Initial Setting; Initial Screen

Endress+Hauser Prothermo NMT81
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Prothermo NMT81
62 Endress+Hauser
Save / Restore
A0044921
 48 Save / Restore screen
Save: The information is sent to a PC from NMT81.
The information of writable parameters regarding device measurements can only be saved on
a PC.
Saving procedure
1. Press [Save / Restore].
2. Check upload from the device before saving in order to save the values that are actually
saved on the device.
3. Press [Save as].
4. Specify the saving destination.
5. Input a file name.
6. Press [Save].
.deh file format is created.
This completes the saving procedure.
Restore: Information saved on a PC is sent to NMT81 again.
Restore procedure
1. Press [Save / Restore].
2. Check [Download data to device after loading].
If data is restored without performing a check, the data is only updated in FieldCare
without sending to the master devices.
The data restore operation without check is used in offline work.
3. Press [Open].

Other manuals for Endress+Hauser Prothermo NMT81

Related product manuals