EasyManua.ls Logo

Endress+Hauser RA33 - 5 Operation; 5.1 General Notes for Operation; 5.2 Display and Operating Elements

Endress+Hauser RA33
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RA33 Wiring
Endress+Hauser 21
4.5.4 Printer interface / RS232 (optional)
The printer/RS232 interface is galvanically isolated (testing voltage: 500 V) and is used to
connect a printer. It is connected via a 3-pin plug-in terminal in the housing cover.
a0014602
Fig. 20: Printer connection via RS232
4.6 Post-connection check
After completing the device's electrical installation, carry out the following checks:
Device status and specifications Notes
Is the device or cable damaged (visual inspection)? -
Electrical connection Notes
Does the supply voltage match the information on the nameplate? 100 to 230 V AC/DC (±10%)
(50/60 Hz)
24 V DC (-50% / +75%)
24 V AC (±50%) 50/60 Hz
Are the mounted cables relieved of tension? -
Are the power supply and signal cables connected correctly? See wiring diagram on the housing

Other manuals for Endress+Hauser RA33