Commissioning Batch Controller RA33
42 Endress+Hauser
Web server settings
MenuÉSetupÉAdvanced setupÉSystemÉEthernetÉWeb serverÉyes or MenuÉExpertÉSys-
temÉEthernetÉWeb serverÉyes
Enter the address for retrieval in the Web browser: http://<IP address>
The following Web browsers are supported:
• MS Internet Explorer 6 and higher
• Mozilla Firefox 2.0 and higher
• Opera 9.x and higher
The operating language for the Web server is English. No other languages are offered.
No provision is made for authentication via ID/password.
Printer interface
The device can print a batch report directly on a connected ASCII printer via the RS232 interface.
a0014552
Fig. 33: Batch controller test report
Menu item Description
Printout You can initiate printing manually on-site if it is set to "Manual“. If it is set to "Automatic",
the configured number of printouts are additionally printed after every batch cycle.
Baudrate Select the baudrate that is compatible with the printer.
Number of copies Use this option to specify the number of printouts for automatic printing at the end of the
batch.
Characters/line Use this option to enter the number of characters per line supported by the printer.
Number of headers Use this option to select the required number of lines for user-defined text at the beginning
of the printout.
Header line 1 to 4 Use this option to enter the free text.
Number of footers Use this option to select the required number of lines for user-defined text at the end of the
printout.
Header line 1 to 4 Use this option to enter the required text.
Blank rows at end Use this option to enter the required blank rows at the end of a printout, e.g. to allow suffi-
cient space for tearing off.
Print direction Use this option to select whether printing should start at the first line or the last line.
Test print Use this option to initiate a test printout.