Loading...
RIA15 Installation
Endress+Hauser 17
2
1
5mm
4mm
1.
2.
3.
4.
TX20
A0017789
 5 Mounting the process display unit on a pipe
1 Mounting plate for pipe/wall mounting
2 Weather protection cover (optional)
1. Release the 4 housing screws
2. Open the housing
3. Secure the mounting plate to the rear of the device with 4 screws supplied. The
optional weather protection cover can be secured between the device and the
mounting plate.
4. Guide the two gripper clamps through the mounting plate, fit them around the pipe
and tighten.
Wall mounting
With optionally available mounting kit.