SPECIFICATIONS
Blanket Warmer ▪ Operat or’s Manu a l ▪ MN -4 6 42 7 ▪ Rev 1 ▪ 6/19
13
SPECIFICATIONS
Specification Information
Transportation and
storage conditions
▪ Ambient temperature range of -40°F to +159°F (-40°C to +70°C).
▪ Relative humidity range of 10% to 95%, non-condensing.
▪ Atmospheric pressure range of 7.25 psi to 15.37 psi (50kPa to 106kPa).
Operating conditions ▪ The warmer must acclimate to the room temperature it will be placed—24 hours
is recommended.
▪ The recommended environmental temperature range is 60°F to 90°F
(15°C to 32°C).
▪ The recommended relative humidity is above 20%, non-condensing.
Standards for
electrical equipment
▪ Medical equipment classified by Underwriters Laboratories with respect
to electrical shock, fire, and mechanical hazards only, in accordance with
UL 61010-1 and CAN/CSA C22.2 No. 61010-1.
▪ Grounding reliability can only be achieved when the appliance is
connected to an equivalent receptacle marked “Hospital Grade.”
▪ IEC 60601-1-2
▪ IP-XO ordinary
Model DC150 DC400
Weight
Net: 56 lb (25 kg)
Ship: 60 lb (27 kg)
Net: 86 lb (39 kg)
Ship: 144 lb (65 kg)
Storage cavity
capacity
3–4 blankets
(1.5 ft
3
)
14–18 blankets
(4.0 ft
3
)
Temperature
range
90°F to 160°F (32°C to 71°C)
Clearance
requirements
3” (76mm) from rear
1” (25mm) from top
1” (25mm) from sides
3/4” (19mm) from bottom
E471516
Protective Earth
Ground Symbol