MAINTENANCE
Blanket Warmer Operat ors Manu a l MN -4 6 42 7 Rev 1 6/19
33
MAINTENANCE
Maintenance Schedule
Daily For daily maintenance, do the following.
Check:
the air vents in the airflow insert panels are not obstructed (if applicable).
all fan guards are clear and not obstructed (if applicable).
the number of bottles/bags, or blankets as applicable do not exceed the
maximum capacity per shelf or basket.
Clean:
any spills with a clean, lint free cloth.
Monthly For monthly maintenance, do the following.
Check:
the door gasket for tears and holes. Make sure that it is firmly attached to the
door. Check the seal when the door is closed.
the guards around the air temperature sensor are in place and fully secure to the
warmer.
the hardware securing the warmer(s) to the wall, if applicable.
Clean:
the outside of the warmer with stainless steel cleaner.
vacuum (if applicable):
fan openings
fan sail switch
vent openings
Yearly For yearly maintenance, do the following.
Check:
the set-point temperature compared to the actual temperature displayed.
Check the cavity air temperature with a thermocouple placed 1” (25mm) from
the cavity sensor. Do not allow the thermocouple to touch any surface.
Monitor the temperature for approximately one hour in an empty cavity.
the condition of the plug and cord and replace if damaged.
the controller screen for excessive wear. Make sure the controller screen displays
and operates properly.
Clean:
the shelves and interior of the warmer.