INSTALLATION
Blanket Warmer ▪ Oper a tor’s Ma nu al ▪ M N- 46 4 2 7 ▪ Rev 1 ▪ 6/19
20
How to Install the Warmer
Before you begin Make sure you have:
An appropriate lifting device and enough personnel to safely move and position
the weight of the warmer.
Requirements
▪ The warmer must be installed on a level surface.
▪ The warmer must not be installed in any area where it may be affected by steam,
dripping water, high temperature, or any other severely adverse conditions.
Voltages
Position the warmer To position the warmer, do the following.
Step Action
1. Make sure that:
▪ The location where the warmer is being installed is rated to support the
weight of the warmer;
▪ The warmer is within five feet of the appropriate electrical outlet;
▪ You follow the warmer clearance guidelines. Refer to topic Specification
Information for clearance requirements.
2. Move the warmer to the installation location and onto the final resting
surface.
3. Lock the casters, if equipped.
The warmer is now correctly positioned.
Continued on next page
Model V Ph Hz kW Plug Configuration
DC150 120
220
230
1
1
1
60
50
50
0.6
0.6
0.6
NEMA 5-15P
CEE 7/7*
BS 1363 (UK only), CEE 7/7*
DC400 120
220
230
1
1
1
60
50
50
0.8
0.8
0.8
NEMA 5-15P
CEE 7/7*
BS 1363 (UK only), CEE 7/7*
*Other international plugs are available, contact the manufacturer for more
information.