Operations Manual
7. CONTROL UNIT OPERATION TYPE 1 AND
TYPE 2
Envirotainer RKN e1 Container
7.2.4 BATTERY LEVEL CHECK
The battery level is shown in the default mode.
The battery percentage figure is continuously updated according to battery status.
100% battery capacity is maximum and is achieved when charging starts from low battery
voltage.
95% battery capacity is considered fully charged if the blue light is fixed (not flashing).
7.2.5 FUNCTIONAL TEST
The recommended procedure is:
1. Disconnect the container from power supply (if connected).
2. Set temperature to +5°C (+41°F).
Fig. 18
3. Check for air discharge from air guide in
the ceiling (Fig. 18 ).
(PROCEED ON NEXT PAGE!)
Check for air
Doc No. MO-RKN-3009
Version 2.2: 2017-08-31 25 (42)