P a g e | 122
6.9.6 Restore All Factory Cal
Tapping “Restore all Factory Cal” will set all field calibration offsets to zero and restore the calibration
values from the last factory service.
If a sensor has drifted significantly, a message indicating such may appear when the restore all factory
cal button is pressed. If restoring factory calibration would adversely affect the ability of the user to
calibrate the unit, the user will be notified and allowed to cancel the operation.