6 Remedying malfunctions
24 / 70
80116-941_V24392
2023-03
APC-A 151 Input pressure < 3 bar (rel)
and bottle pressure
> 25 bar (rel).
[Vacuum in the gas supply,
gas hose not connected]
Check the pressure regu-
lator and gas bottle. If a
gas bottle is not used,
check the pressure of the
central gas supply.
If the error occurs again,
notify Erbe Service.
Display on the unit: APC or APC-A, detecting module: GCM