4 Technical Data
80116-941_V24392
2023-03
CHAPTER 4
Technical Data
Connections
Low voltage Via VIO 3 electrosurgical unit
HF Via VIO 3 electrosurgical unit
Grounding terminal yes
Gas specification
Type of gas Argon
Argon minimum purity Argon 4.8 (99.998% purity) or higher, e.g. argon 5.0.
Density (relative; air = 1) 1.38
Critical temperature –122 °C
Color Colorless gas
Odor No warning by smell
Concentration of explosion limit (vol. % in air) Non-combustible
Special risks Heat/fire can result in an increase in pressure caus-
ing the pressure gas bottle to explode! Gas in high
concentrations can be suffocating!
Gas-specific unit data
Input pressure (5 ±2) x 10
5
Pa 5 ±2 bar 72.5 ±29 psi
Max. output pressure 2 x 10
5
±2 x 10
4
Pa 2 ±0.2 bar 29 ±2.9 psi
Adjustable gas flow 0.1 - 8 l/min limited by the respective instrument
connected, settable in 0.1 l steps
Tolerance of the nominal flow ±20% (in the range 0.1 - 8 l/min)
Purge flow Instrument-dependent (corresponds to the setpoint
flow, which is currently connected)
Purge time Factory setting: 3 sec.