7 • Maintenance and servicing
80116-941_V24392
2023-03
Pressures
Testing and measuring equipment
Vacuum/overpressure
detection
Test setup
Fig. 7-1
IMPORTANT! For this test only Erbe pressure reducers may be used. For article num-
bers see chapter "Testing and measuring equipment".
Erbe Art. No. Description
20100-108 Manometer
20134-XXX
1
1. dependent on bottle connector.
Pressure regulator
20100-XXX
1
Variable pressure regulator
– Argon gas bottle
20100-175 FiAPC test adapter
20194-070 Patient cable NE
or
20194-075 Patient cable NE, international
20100-033 Adapter cable NE
20189-350 Dual-pedal footswitch with ReMode