7 • Maintenance and servicing
42 / 70
80116-941_V24392
2023-03
Test procedure
1. Set the test specimen to:
APC probe, monopolar – forcedAPC – Effect 0.1
2. Set argon flow to 0.1 l/min.
To do so, switch to the "Effect Setting" window and press <Argon flow
l/min>.