21 Generator G11
Generator G11 Operator’s Manual
The instrument may not be properly
assembled to the hand piece�
Press the ‘Next’ button to continue�
Re-tighten the instrument� Ensure that one hand
is holding the appropriate torque wrench and
the other hand is holding the hand piece (not the
instrument)�
If using shears, keep jaws of instrument
open during test�
Press ‘Next’ to advance, to retry test with jaws
open�
If using shears, keep jaws of instrument
open during test�
Press ‘Next’ to advance, to retry test with jaws
open�
Indicates an internal error with the
instrument that prevents it from being
used�
Unplug the hand piece and replace the
instrument�
Note: If multiple instruments have indicated
an instrument error on the same hand piece, try
replacing the hand piece to ensure that the issue
is related to the instruments�
A system reset is required� Reactivate instrument to reset system and
continue�
Indicates an internal error with the
identification circuitry in the instrument�
Special circuitry is installed on certain
advanced instruments to identify the device
to the generator and tailor the generators
output to the instrument�
Unplug hand piece and replace instrument�