27 Generator G11
Generator G11 Operator’s Manual
Relax Pressure on Blade�
(Note: Press ‘Settings’ for Settings Menu�)
Clean Blade� Remove any tissue which may be lodged inside end of instrument sheath�
(Note: Press ‘Next’ to advance�)
Activate instrument outside patient to run test� If using shears, open jaws during test�
HARMONIC Settings:
Display Brightness
Hand / Foot Activation
Language
Tone Change ON/OFF
Convert MIN to Adv Hemostasis
HARMONIC Settings:
Convert MIN to Adv Hemostasis
Hand Piece Test
System Information
Biomed Information
HARMONIC Display Brightness�