Eversense CGM User Guide
Q: Why do I see a “New Sensor Detected” notification?
A: This message appears when your smart transmitter
detects a new sensor so you may link the smart
transmitter and sensor. The smart transmitter can only
be linked to one sensor at a time. If you see a New
Sensor Detected message and you already have
a sensor inserted and linked to your smart transmitter,
tap Not Now. If unsure, contact Customer Support for
more information.
Q: Why did my CGM System re-enter Initialization Phase?
A: You will re-enter Initialization Phase for any one of the
following reasons:
Calibration period has expired without you having
entered a fingerstick test value.
3 or more blood glucose readings are significantly
dierent than the current sensor glucose readings.
Y our smart transmitter has not been charged within
16 hours of the empty battery alert.
If you manually change the time on your mobile
device your smart transmitter will sync and reinitialize
to your mobile device.
Q: Is it okay for an MRI technician to wear the
Eversense CGM System?
A: Yes, MRI technicians can wear the Eversense
CGM System. However, for people undergoing an
MRI with a static magnetic field of 1.5T or 3.0T,
the sensor can stay in place under the skin, but
the smart transmitter must be removed and left
outside the room.
LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 166LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 166 2/26/20 12:59 PM2/26/20 12:59 PM