Alert Setting
(mg/dL)
Confirmed
Event Detection
Rate
Missed
Detection
Rate
True Alert
Rate
False Alert
Rate
Low
Alert
60 76% 24% 81% 19%
70 90% 10% 95% 5%
80 94% 6% 96% 4%
90 96% 4% 95% 5%
High
Alert
120 99% 1% 96% 4%
140 99% 1% 96% 4%
180 98% 2% 95% 5%
200 98% 2% 95% 5%
220 98% 2% 95% 5%
240 98% 2% 95% 5%
300 93% 7% 94% 6%
Table 11 – Eversense High and Low Glucose Alert Performance (Threshold Only) in the PRECISION Study
The table below shows an assessment of the ability of the Eversense CGM System to detect high and low glucose
levels, and assess true positive alerts vs false positive alerts. These are shown as a percentage of alerts provided
compared to the YSI reference values at various thresholds.