19
185
Eversense CGM User Guide
Table 15 – Concurrence with YSI Range Day 1 in the PRECISION Study
Percent of Matched Pairs in Each YSI Glucose Range for Each CGM Glucose Range
YSI (mg/dL)
CGM
(mg/dL)
Paired
CGM-YSI
(n)
< 40 40-60 61-80 81-120 121-160 161-200 201-250 251-300 301-350 351-400 > 400
40 - 60 274 1% 59% 36% 4% 0% 0% 0% 0% 0% 0% 0%
61 - 80 378 0% 14% 65% 19% 1% 0% 0% 0% 0% 0% 0%
81 - 120 516 0% 1% 23% 63% 12% 0% 0% 0% 0% 0% 0%
121 - 160 332 0% 0% 0% 17% 70% 12% 1% 0% 0% 0% 0%
161 - 200 226 0% 0% 0% 1% 23% 58% 17% 0% 0% 0% 0%
201 - 250 268 0% 0% 0% 0% 4% 25% 55% 15% 0% 0% 0%
251 - 300 205 0% 0% 0% 0% 0% 2% 25% 57% 14% 1% 0%
301 - 350 250 0% 0% 0% 0% 0% 0% 4% 27% 42% 26% 1%
351 - 400 216 0% 0% 0% 0% 0% 0% 1% 9% 46% 40% 4%
LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 185LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 185 2/26/20 12:59 PM2/26/20 12:59 PM