19
189
Eversense CGM User Guide
Days 1 - 30 Days 31 - 60 Days 61 - 90 All Days
Number of
Displayed
Glucose Values
279,674 274,727 266,895 821,296
Expected
Number of
Glucose Values
285,579 280,478 276,757 842,814
% of Displayed
Glucose Values
98% 98% 96% 97%
Data Availability
A total of 97% (821,296 out of expected 842,814) glucose values were available during the 90 day study.
Table 17 – Eversense Data Availability in the PRECISION Study