Evoqua IONPURE IP-POWERDSP-G2 - User Manual

Evoqua IONPURE IP-POWERDSP-G2
25 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
10 Technology Drive
Lowell, Massachusetts 01851
Tel: (866) 876-3340 Fax: 978-934-9499
www.ionpure.com
IONPURE
®
G2 DISPLAY
BOARD
For use with G2
Power Controllers
Operation
&
Maintenance
Manual
IP-POWERDSP-G2MAN
November 2011
Revision 1
Manual Covers Part #:
IP-POWERDSP-G2

Questions and Answers

  • C
    choirichardAug 9, 2025
    What to do if my Evoqua Control Panel displays 'ADDRESS CONFLICT'?
    • S
      serranoearlAug 9, 2025
      If an Evoqua Control Panel channel displays 'ADDRESS CONFLICT', it means that two or more power controllers are set to the same SW2 position. To resolve this, set a unique ID to each power controller. Another possible cause is that the display and power controller have incompatible firmware revisions; in this case, contact your parts supplier to update the firmware.
  • W
    williamskarenAug 15, 2025
    Why is a channel row blinking on my Evoqua Control Panel?
    • M
      Malik BeltranAug 15, 2025
      If a channel row is blinking on your Evoqua Control Panel, it indicates that an alarm has triggered. Access the ALARMS submenu to identify the specific alarm that has been activated.
  • T
    Tina GutierrezAug 19, 2025
    How to fix asterisks on an Evoqua Control Panel channel?
    • N
      Nathan WilliamsAug 19, 2025
      If an Evoqua Control Panel channel displays asterisks, it could be due to several reasons. First, verify that the Ethernet cable is functioning correctly and is not a crossover cable. Second, ensure the power controller in that channel is energized. Third, check that the SW2 position is set to the correct channel using Table 1. Finally, the display and power controller might have incompatible firmware versions, requiring you to contact your parts supplier for a firmware update.
  • D
    Douglas JohnsonAug 25, 2025
    How to troubleshoot an Evoqua Control Panel display that does not power on?
    • S
      Savannah SummersAug 25, 2025
      If the display of your Evoqua Control Panel does not power on, there could be a couple of reasons. First, check the Ethernet cable to ensure it is working correctly and is not a crossover cable. Second, confirm that the power controller is energized.
  • V
    victoria25Aug 29, 2025
    What to do if I can’t adjust the DC output current past a certain value on my Evoqua Control Panel?
    • M
      mariarosarioAug 29, 2025
      If you can't adjust the DC output current past a certain value on your Evoqua Control Panel, it could be because the current range is set too low on the DC power controller. Select the adequate current range for the CEDI module. Alternatively, the power controller may have reached its maximum DC output, so verify that the input voltage to the power controller is adequate and that the resistance of the CEDI module is within specifications. Finally, the “VDC limit” might be set too low; access the SETUP submenu to adjust the “VDC limit” value.
  • J
    jeremymarshallSep 3, 2025
    What to do if Evoqua Controller indicates input under-voltage?
    • A
      allentroySep 3, 2025
      If the AC input voltage to your Evoqua Controller G2 drops below 726 V, the DC power stage will restart. However, if this happens three times within 30 minutes, the DC power stage will shut down. To restart it, you'll need to manually cycle the AC input power to the controller off and then on again.