IONPURE
®
G2 DISPLAY BOARD – IP-POWERDSP-G2
IP-POWERDSP-G2MAN Rev 1 Page 18 of 25
Change password?: press ENTER and use the UP/DOWN keys to change the
password of the SETUP submenu.
PAGE 2:
Figure 16: SETUP Screen – Page 2
Change module?: it is recommended to use this option to reset all historic data
after a module replacement. Pressing ENTER will reset the following:
- Op.hours value in RUN submenu
- Ohms init value in RUN submenu
- Resistance alarm in ALARMS submenu
- All logs in the LOGS submenu
ALARMS: this submenu displays the status of alarms and provides the option to clear them.
There are 2 pages in this submenu. Data is grouped as follows:
PAGE 1:
Figure 17: ALARMS Screen – Page 1