Chromatic Dispersion Analyzer 5
2 Getting Started with Your
Chromatic Dispersion
Analyzer
Inserting and Removing Test Modules
To insert a module into the FTB-400 Universal Test System:
1. Exit ToolBox and turn off your unit.
2. Position the FTB-400 so that its right panel is facing you.
3. Take the module and place it so that the connector pins are at the
back, as explained and shown below. Identification sticker must be
facing down and connector pins at the left of the retaining screw hole.
IMPORTANT
The power consumption of the module is relatively high. Therefore,
it is not recommended to use it with battery or to activate other
power consuming modules at the same time on the platform.
CAUTION
Never insert or remove a module while the FTB-400 Universal Test
System is turned on. This will result in immediate and irreparable
damage to both the module and unit.
WARNING
When the laser safety LED ( ) is flashing on the FTB-400, at least
one of your modules is emitting an optical signal. Please check all
modules, as it might not be the one you are currently using.
CAUTION
Inserting a module upside down could result in permanent damage
to the module, as the connector pins might be bent.