Test Setup - Test Configurator, Timer, and System
120 MAX-800 Series
Labels
Labels
From the Test menu, tap Setup, Test Configurator, the interface block,
and on the Labels tab.
Note: Selecting a Label byte to be generated will automatically update the
corresponding OH byte. Refer to OH - SONET/SDH on page 343 for more
information.
Labels
STS/AU Path (C2): The C2 byte is allocated to indicate the content of
the STS SPE / VC, including the status of the mapped payloads.
Generated: Select the C2 byte from the list. Selecting a C2 byte value
will automatically update the OH C2 selection and vice versa. Refer to
C2 on page 351 for more information.
PLM-P/UNEQ-P / HP-PLM/HP-UNEQ: Enables the Payload Mismatch
and STS/AU UNEQ monitoring. This setting is coupled with the
configuration of Traces (SONET/SDH) on page 301.
Expected: Select the expected C2 byte from the list. Refer to C2 on
page 351 for more information.
VT/TU Path (V5): The V5 byte is allocated to indicate the content of the
VT/TU path, including the status of the mapped payloads.
Generated: Select the V5 byte from the list. Selecting a V5 byte value
will automatically update the OH V5 selection and vice versa. Refer to
V5 on page 354 for more information.
PLM-V/UNEQ-V / LP-PLM/LP-UNEQ: Enables the Payload Mismatch
and VT/TU UNEQ monitoring. This setting is coupled with the
configuration on the Labels on page 259.
Expected: Select the expected V5 byte from the list. Refer to V5 on
page 354 for more information.