Test Functions
322 MAX-800 Series
BFD (Bidirectional Forwarding Detection)
BFD (Bidirectional Forwarding Detection)
Note: Only available with IPv4 Network Layer (refer to IP on page 134) in single
port topology.
From the Test menu, tap Functions, and the BFD tab.
Bidirectional Forwarding Detection
Local IP Address displays the configured interface Source IP Address
(refer to IP on page 134).
Remote IP Address allows configuring the destination IP Address.
IP TOS/DS allows entering the type of service: 00 (default) to FF.
Desired Min TX Interval defines the minimum interval the local unit
would like to use when transmitting BFD control packets:
100 ms (default) to 1 s.
Required Min RX Interval defines the minimum interval between
received BFD control packets the local unit requires: 100 ms (default)
to 1 s.
Detect Multiplier is a value that is multiplied by the negotiated
transmit interval giving the detection time for the receiving system.
Choices are 1 to 255 (3 by default).
Session State indicates the local session state: Down, Init, Up.
Diagnostic indicates the reason of the last local session state change.
Connect starts/stops the BFD process.
Statistics
Packets Transmitted indicates the number of sent BFD control
packets.
Packets Received indicates the number of received BFD control
packets.