11.0 MAINTENANCE
This section has a selection of special utility programs for special
functions
11.1 MAINTENANCE MENU
This is the main Maintenance menu – see fig. 36
Fig. 36
11.2 LIBRARY
Selection of this option displays the nuclides in the system library
– see Fig. 37. For the current library list refer to Appendix D
Fig. 37
11.3 CHARGING
This special feature is used during system setup
IT IS NOT RECOMMENDED FOR GENERAL USE
AS IT CONTAINS NO USEFUL INFORMATION
FOR THE NORMAL USER.
Fig. 38
11.4 REMOTE
This sets the RS232 port to the active state to permit data transfer
using the Exploranium IdentiView software described in Appendix B.
This setting works as a latch – the first activation sets the port
ACTIVE and the next activation sets it INACTIVE. Normally the
port should be INACTIVE for normal system functions.
In the system architecture the RS232 port and the AUDIO sound
share a common data point. This is normally not an issue as the
Audio is not normally active during RS232 functions.
Users are advised that when the REMOTE mode is active, the AUDIO is disabled.
CHARGING
BAT VOLTAGE 243
EL.METER 01273
EXT.POWER OFF
CHARGING OFF
CH.TIME 00000
REFRESH OFF
ERR.STATUS 02
LIBRARY
CHARGING
REMOTE
DEFAULT
ADC TEST
MAIN MENU
MAINTENANCE
STANDARD LIBRARY
U-238 RA-226
TH-232 K-40
CE-139 I-131
CS-137 CO-60
BA-133 TC99M
CO-57 AM-241
XE-133 NP-237
PD-103 IR-192
ENTER TO GET MORE
LIBRARY
LIBRARY
CHARGING
REMOTE
DEFAULT
ADC TEST
MAIN MENU
MAINTENANCE
REMOTE
MODE
ACTIVE
MESSAGE