Colon Insufflator User Manual
Page
30
Condition
Possible Causes
Field Checkout
“CONTINUED”
My
PROTOCO
2
L Colon
Insufflator
does not deliver
CO
2
.
What could be the problem?
PROTOCO
2
L Colon
Insufflator
flow rate
incorrect
At start
-
up, your
P
ROTOCO
2
L Colon Insufflator
should
normally deliver the flow of CO
2
to the patient incrementally
as follows: 1 LPM for 0 to 0.5 Liters of delivered volume, 2
LPM for greater than 0.5 to 1 Liter of delivered volume and
a maximum of 3 LPM for delivered volu
mes greater than 1
Liter
.
This procedure verifies this performance parameter. If
you are unable to successfully replicate these steps, contact
E
-
Z
-
EM, Inc. Customer Solutions or check the routing of
patient Administration Set to make sure that it is
not
ki
nked,
occluded or leaking.