Colon Insufflator User Manual
What could be the problem?
normally deliver the flow of CO
to the patient incrementally
as follows: 1 LPM for 0 to 0.5 Liters of delivered volume, 2
LPM for greater than 0.5 to 1 Liter of delivered volume and
a maximum of 3 LPM for delivered volu
This procedure verifies this performance parameter. If
you are unable to successfully replicate these steps, contact
EM, Inc. Customer Solutions or check the routing of
patient Administration Set to make sure that it is