Colon Insufflator User Manual
What could be the problem?
left open all of the time.
We recommend attaching your
accessory to either “D” or “E” size CO
tank. Be sure to close the CO
supply tank valve whenever
The pneumatic design of the
is not intended to provide a full time gas seal
supply tank while the device is not being used or in
supply tank valve inadvertently be
left open, there is a high likelihood that the CO
contents will gradually empty over several days.
your procedure schedule, you still encounter problems with
supply tank prematurely emptying, there is the
possibility that the High Pressure Hose from the CO
Please inspect the High Pressure Hose in accordance with the
following procedure: Using an adjustable or Open Ended ¾"
wrench, detach the High Pressure Hose from
’s rear panel. Inspect the
apered surface of the free standing nipple on the
High Pressure Hose. This is the sealing surface. Verify that it
is free of any debris or contaminant. Also, verify that the
surface is smooth and free of
Repeat this inspection procedure at the identical tapered gas