Loading...
PROTOCO
2
L
Colon Insufflator User Manual
Page
34
Condition
Possible Causes
Field Checkout
“CONTINUED”
My CO
2
supply
Tank
is
prematurely emptying.
What
could be the problem?
High Pressure Hose from
CO
2
supply tank is not
installed properly
Inspect the CO
2
supply tank Yoke and verify that the plastic
disc that creates the gas seal at the CO
2
supply tank valve is
in
-
place and free from defect.
Af
ter verifying all High Pressure Hose sealing surfaces are
clean and free of defect; reconnect the High Pressure Hose to
the CO
2
supply tank Yoke and
PROTOCO
2
L Colon
Insufflator
. Use an adjustable or Opened Ended ¾" wrench
to tighten securely.
DO NOT USE A
NY TEFLON TAPE
OR THREAD SEALING COMPOUNDS
.
Reconnect the Yoke to the CO
2
supply tank valve making
sure that the plastic sealing disk is in place.
Open the CO
2
supply tank valve. If
CO
2
continues to
prematurely empty
or should you hear CO
2
escaping from
an
y of these connections or the High Pressure Hose itself,
contact E
-
Z
-
EM, Inc. Customer Solutions
.
Disc