FEMA ELECTRÓNICA . Series C . C40-D
4
Digits
number of digits 4
led 7 segments led
color red
height 14 mm
Reading
max. reading 9999
min. reading -1999
decimal point congurable X.X.X.X
readings 3 readings / second
display refresh 3 refresh / second
step response 300 mSec. (0 % to 99 % signal)
overrange reading ashes at ‘9999’
underrange reading ashes at ‘-1999’
Accepted input signals
AC voltages and currents 600 Vac, 200 Vac, 20 Vac, 2 Vac
200 mVac, 60 mVac, 5 Aac, 20 mAac
(see secon 1.10)
DC voltages and currents ±600 Vdc, ±200 Vdc, ±20 Vdc, ±2 Vdc
±200 mVdc, ±60 mVdc, ±5 Adc, ±20 mAdc
(see secon 1.11)
thermocouples K, J, E, N, L, R, S, B, T y C
(see secon 1.12)
temperature ‘Pt’ Pt100 (2 and 3 wires with automac
compensaon up to 30 R), Pt500, Pt1000
(see secon 1.13)
temperature ‘Ni’ Ni100, Ni200, Ni1000
(see secon 1.13)
temperature ‘NTC’ (see secon 1.14)
temperature ‘PTC’ (see secon 1.15)
process 4/20 mA, 0/10 Vdc (acve and passive)
(see secon 1.16)
frequency up to 100 Hz
(see secon 1.17)
resistances ranges of 5 K and 50 K
(see secon 1.18)
potenometers nominal value 500 Ohm up to 20 KOhm
(see secon 1.19)
Accuracy at 25 ºC see following secons for each signal
Thermal dri 150 ppm/º
Excitaon voltage +15 Vdc (max. 30 mA) for process signals
+5 Vdc for potenometers
(at terminal 5, see secons 1.5 and 1.16)
Power supply
power ‘U’ 18 to 265 Vac/dc
isolaon 1500 Ve.
isolaon tested for 60 sec.
AC frequency 50/60 Hz
consumpon
(without ‘Eco’)
<1.5 W meter only
<2.5 W meter with opons
consumpon
(with ‘Eco’)
<0.3 W meter only
<1.5 W meter with opons
Power wires 1 mm
2
to 2.5 mm
2
(AWG17 to AWG14)
Conguraon 3 front push buons
Front protecon IP50 standard
IP65 oponal (see secon 3.2)
Output opons relay, analog, serial
(see secon 2)
Mechanical
mounng panel
connecons plug-in screw terminals
housing material ABS, polycarbonate (V0)
weight <150 grams
front size 96 x 48 mm (1/8 DIN)
panel cut-out 92 x 44 mm
deep 91 mm (including terminals)
Temperature
operaon from 0 to +50 ºC
storage from -20 to +70 ºC
‘warm-up’ me 15 minutes
1.8 Technical specicaons
Funcons included Secon
Fast access to alarm setpoints, maxi-
mum and minimum
1.20
External control second scaling
decimal point 0, 1, 2 or 3
‘hold’ reading
tare funcon
memory of maximum
memory of minimum
1.24
‘Eco’ mode reduced consumpon 1.23
Alarms setpoint
hysteresis
set as max or min type
1.29.3
Oset reading subtract a xed number
of counts to reading
1.22
Display lter recursive
‘steps’
1.29.7
Simplied scaling 1.21
Memory max and min memory 1.29.4
Password blocks conguraon 1.29.7
Display brightness 5 levels 1.29.7
Table 1 - Funcons included