FLUXUS ADM 8027, F801, ADM 8127B A Menu Structure
142 UMFLUXUS_F8V4-6-1EN, 2018-10-10
input of the cut-off flow for negative measured values
range: -12.7...0 cm/s
default: -2.5 cm/s
This display will only be indicated if Cut-off Flow = sign and Cut-off
Flow = user is selected.
x
input of the cut-off flow for the absolute value of the measured values
range: 0…12.7 cm/s
default: 2.5 cm/s
This display will only be indicated if Cut-off Flow = absolut and Cut-
off Flow = user is selected.
x
input of the max. signal amplification
range: 0...255
0: no limit of the signal amplification
This display will only be indicated if the SuperUser mode is activated.
x
input of the fixed upper limit of the sound speed
range: 0...3 000 m/s
0: the default limit of 1 848 m/s will be used
This display will only be indicated if the SuperUser mode is activated.
x
Input of the offset.
range: 0...900 m/s
0: the default limit of 300 m/s will be used
This display will only be indicated if the SuperUser mode is activated.
x
activation of the totalizers overflow x
activation of the taking-over of the totalizer values after the measurement re-
start
x
input of a time interval after which the transmitter has to detect a long mea-
surement failure if there are no valid measured values
0: the default value of 30 s is used
This display will only be indicated if the SuperUser mode is activated.
x
input of the number of decimal places of the totalizers
Automatic: dynamic adjustment
Fixed to x digit: 0...4 decimal places
This display will only be indicated if the SuperUser mode is activated.
x
activation of the manual reset of the totalizers
This display will only be indicated if the SuperUser mode is activated.
x
activation of the display of the totalizer sum
This display will only be indicated if the SuperUser mode is activated.
x
activation of the display of the last valid measured value
This display will only be indicated if the SuperUser mode is activated.
x
activation of the turbulence mode x
INIT-resis-
tant
-Cut-off Flow
-2.5 cm/s
Cut-off Flow
2.5 cm/s
A: Gain threshold
Fail if> 90
A: Bad soundspeed
thresh. 2007 m/s
A: Bad soundspeed
offset: +321
Quant. wrapping
off >ON<
Quantity recall
off >ON<
Do not total. if
no meas.> 0 s
Total digits
Automatic
3xC clear totals
off >ON<
ShowΣQ
off >ON<
Keep display val
off >ON<
Turbulence mode
off >ON<

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS ADM 8027 and is the answer not in the manual?

Flexim FLUXUS ADM 8027 Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS ADM 8027
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals