7 Specifications
A6700sc/A6750sc User’s Manual
41
Power Consumption
Camera DC Power @ 24V
DC
Continuous Cool Down: 24 Watts
Continuous Normal: 21.25 Watts
Cool-down Time
≈7 minutes to reach operating temperature
Sensitivity (w/o optics)
NEΔT
1
18 mK typ.
1) NET is at 50% nominal bucket fill, 298K background, + 5
o
C signal
7.6 Non Uniformity Correction
NUC Types
One Point (offset value with unity gain)
Two Point (offset and gain values) non-volatile
Two Point w/Bad Pixel Detection/Replacement
Update Offset (recalculates offset using current gain)
NUC Source
Internal: Ambient flag (for 1-pt and offset update only)
External: Any user supplied source which covers entire FOV
Bad Pixel Replacement
Two-Point Gradient and nearest neighbor (autoselected)
Number of NUC’s
4 active NUC’s in preset selectable form
>100 NUCs in on-board flash
NUC Time
< 15 seconds
NUC Performance
0.1%
7.7 Detector/FPA
Spectral Response
3-5um (1-5um for broadband models)
Detector Type
InSb
f/#
f/2.5 or f/4
Integration Mode
Snapshot
Format (HxV)
640 x 512
Operability
>99.8%, 99.95% typical
Charge Handling
Capacity
1
7.2 x10
6
e-
Detector Pitch
15 microns
Detector Cooling
Rotary Cryo Cooler