EasyManuals Logo

FLIR E40 User Manual

FLIR E40
183 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #142 background imageLoading...
Page #142 background image
Cleaning the camera
26
26.1 Camera housing, cables, and other items
26.1.1 Liquids
Use one of these liquids:
Warm water
A weak detergent solution
26.1.2 Equipment
A soft cloth
26.1.3 Procedure
Follow this procedure:
1. Soak the cloth in the liquid.
2. Twist the cloth to remove excess liquid.
3. Clean the part with the cloth.
CAUTION
Do not apply solvents or similar liquids to the camera, the cables, or other items. This can cause
damage.
26.2 Infrared lens
26.2.1 Liquids
Use one of these liquids:
A commercial lens cleaning liquid with more than 30% isopropyl alcohol.
96% ethyl alcohol (C
2
H
5
OH).
26.2.2 Equipment
Cotton wool
26.2.3 Procedure
Follow this procedure:
1. Soak the cotton wool in the liquid.
2. Twist the cotton wool to remove excess liquid.
3. Clean the lens one time only and discard the cotton wool.
WARNING
Make sure that you read all applicable MSDS (Material Safety Data Sheets) and warning labels on con-
tainers before you use a liquid: the liquids can be dangerous.
CAUTION
Be careful when you clean the infrared lens. The lens has a delicate anti-reflective coating.
Do not clean the infrared lens too vigorously. This can damage the anti-reflective coating.
#T559845; r. AJ/37554/37554; en-US
132

Table of Contents

Other manuals for FLIR E40

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the FLIR E40 and is the answer not in the manual?

FLIR E40 Specifications

General IconGeneral
Resolution160 x 120 pixels
Accuracy±2°C or ±2% of reading
FocusManual
Image StorageSD card
Image Frequency60 Hz
Digital Camera3.1 MP
Battery operating timeApprox. 4 hours
Operating Temperature-15°C to +50°C
Storage Temperature-40°C to +70°C
Spectral Range7.5 - 13 µm
Field of View25° x 19°
Display3.5 in color LCD
Image ModesThermal, Visual, MSX
Measurement PresetsCenter Spot, Hot Spot, Cold Spot
File formatJPEG
Battery typeLi-Ion battery
ConnectivityUSB, Wi-Fi
Thermal Sensitivity< 0.07°C @ +30°C (+86°F) / 70 mK