EasyManuals Logo

FLIR E40 User Manual

FLIR E40
183 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #154 background imageLoading...
Page #154 background image
Thermographic measurement techniques30
30.2.1.1.1 Method 1: Direct method
Follow this procedure:
1. Look for possible reflection sources, considering that the incident angle = reflection
angle (a = b).
Figure 30.1 1 = Reflection source
2. If the reflection source is a spot source, modify the source by obstructing it using a
piece if cardboard.
Figure 30.2 1 = Reflection source
#T559845; r. AJ/37554/37554; en-US
144

Table of Contents

Other manuals for FLIR E40

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the FLIR E40 and is the answer not in the manual?

FLIR E40 Specifications

General IconGeneral
Resolution160 x 120 pixels
Accuracy±2°C or ±2% of reading
FocusManual
Image StorageSD card
Image Frequency60 Hz
Digital Camera3.1 MP
Battery operating timeApprox. 4 hours
Operating Temperature-15°C to +50°C
Storage Temperature-40°C to +70°C
Spectral Range7.5 - 13 µm
Field of View25° x 19°
Display3.5 in color LCD
Image ModesThermal, Visual, MSX
Measurement PresetsCenter Spot, Hot Spot, Cold Spot
File formatJPEG
Battery typeLi-Ion battery
ConnectivityUSB, Wi-Fi
Thermal Sensitivity< 0.07°C @ +30°C (+86°F) / 70 mK