FLIR Fido® X4 Operations Manual
Step 4: The TD slot will automatically close and the message “Checking System Per-
formance” will be displayed along with a progress indicator.
Figure 5-7: Checking System Performance screen
Caution: The metal portions of the reference standard can be
HOT. Handle with care to avoid burns.
Step 5: When the TD slot opens, remove the reference standard.
Step 6: Close the reference standard by lightly pressing the thumb button and sliding
until the extension is fully retracted back into the sheath.
Step 7: Store the closed reference standard in the kit.
Step 8: When analysis is complete, the Fido X4 will automatically display the result:
“Verify Pass," "Verify Fail," or an error message will appear. If “Verify Pass
appears (Figure 5-8), select OK.
Figure 5-8: Verify Pass screen