EasyManua.ls Logo

FLIR Fido X4 - Replacing the Detector Inlet

FLIR Fido X4
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Step 2: Rotate the TD up to click the latches into place.
8.5 Replacing the Detector Inlet
The Fido X4 is equipped with a user-replaceable detector inlet (3.4.2 , p.17). The inlet should
only need to be replaced in the event of a serious clog or physical damage. A spare detector
inlet and additional O-rings are provided in the maintenance kit. The inlet can also be removed
for cleaning.
8.5.1 Removing the inlet
Step 1: Power OFF the Fido X4 and allow 15 minutes for surfaces to cool.
Step 2: Remove the sensing element (see 8.1.1).
Step 3: Using the wrench component of the multi-purpose inlet tool provided in the
maintenance kit, loosen the inlet by turning counterclockwise.
Figure 8-9: Inlet tool
Step 4: There are O-rings located inside of the inlet. Ensure they are accounted for
before proceeding.
Step 5: Place the inlet on a clean surface.
8.5.2 Installing the inlet
Step 1: Ensure the O-rings are installed inside of the inlet.
Step 2: Thread the inlet onto the system by-hand by turning clockwise.
Step 3: Use the inlet tool to tighten.
Step 4: Do not over-tighten. The inlet will come to a hard stop when seated correctly.
Step 5: Power on the system and re-install a sensing element.
4226044 (Rev. C) © 2020 FLIR Systems, Inc.
59
FLIR Fido® X4 Operations Manual

Table of Contents