12.3 Infrared detector
General
Even small amounts of dust on the infrared detector can result in major blemishes
in the image. To remove any dust from the detector, follow the procedure below.
NOTE
This section only applies to cameras where removing the lens exposes the infrared
detector.
In some cases the dust cannot be removed by following this procedure: the infrared
detector must be cleaned mechanically. This mechanical cleaning must be carried
out by an authorized service partner.
CAUTION
In Step 2 below, do not use pressurized air from pneumatic air circuits in a workshop,
etc., as this air usually contains oil mist to lubricate pneumatic tools.
Procedure
Follow this procedure:
Remove the lens from the camera.1
Use pressurized air from a compressed air canister to blow off the dust.2
Publ. No. T559580 Rev. a486 ENGLISH (EN) November 17, 2010 37
12 Cleaning the camera

Table of Contents

Other manuals for FLIR i3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the FLIR i3 and is the answer not in the manual?

FLIR i3 Specifications

General IconGeneral
Infrared Resolution60 x 60 pixels
Thermal Sensitivity< 0.15°C
Temperature Range-20°C to +250°C
Field of View12.5° x 12.5°
FocusFixed
Battery Life5 hours
Operating Temperature-10°C to +50°C
Storage Temperature-40°C to +70°C
Display2.8 in color LCD
Weight0.34 kg