K
kathrynsmithAug 3, 2025
What to do if my Flowserve Valve Positioners position is driven fully open or closed?
- AAbigail ThomasAug 3, 2025
If the position is driven fully open or closed and won't respond to commands, here's what you can check: 1. Perform stroke calibration (Quick-Cal). 2. Verify hardware connections of the inner-loop hall sensor. 3. Check ATO (Air-to-open) and ATC (Air-to-Close) settings. Recalibrate using Quick-Cal to apply settings. 4. Verify ATO/ATC actuator tubing. 5. Replace the electro-pneumatic converter, if it is malfunctioning. 6. Adjust inner-loop if the control parameter inner-loop offset is too high or low and see if proper control resumes.