Models 110, 111 & 112
Calibration Information
12
Table 2. Performance Test Steps
Display
Function
(switch position) Input value
Frequency
(sine wave)
Lower Limit Upper Limit
300 mV 50 Hz 0.294 0.306
300 mV 500 Hz 0.294 0.306
5 V 50 Hz 4.947 5.053
5 V 500 Hz 4.947 5.053
50 V 500 Hz 49.47 50.53
B
600 V 500 Hz 593.7 606.3
B + Hz
5 V 50 kHz 49.93 50.07
0 V -0.002 0.002
5 V 4.994 5.006
-50 V -49.94 -50.06
D
600 V
0 Hz
599.4 600.6
20 Beeper On
s
250
N/A
Beeper Off
0 -0.2 0.2
1
500 495.3 504.7
1
5 k 4.954 5.046
50 k 49.54 50.46
500 k 495.4 504.6
5 M 4.954 5.046
e
30 M
N/A
29.52 30.48
R
2 V 0 Hz 1.980 2.020
S
0.8 µF N/A 783 nF 817 nF
J
(Models 111 & 112)
9 A 500 Hz 8.83 9.17
5 A 4.047 5.053
-5 A -4.047 -5.053
I
(Models 111 & 112)
9 A
0 Hz
8.88 9.12
1 Does not include test lead resistance.

Other manuals for Fluke 112

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 112 and is the answer not in the manual?

Fluke 112 Specifications

General IconGeneral
CategoryMultimeter
ContinuityYes
Diode TestYes
DisplayDigital
DC Voltage Ranges400.0 mV, 4.000 V, 40.00 V, 400.0 V, 600 V
Resistance Ranges400.0 Ω, 4.000 kΩ, 40.00 kΩ, 400.0 kΩ, 4.000 MΩ, 40.00 MΩ
Safety CategoryCAT III 600 V
AC Voltage Accuracy±(1.0% + 3)
Battery Type9 V
DC Voltage Accuracy±(0.5% + 2 counts)
AC Voltage Ranges4.000 V, 40.00 V, 400.0 V, 600 V