Models 110, 111 & 112
Calibration Information
4
Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C (64 °F to 82 °F), with
relative humidity at 0 % to 95 %.
The accuracy specifications take the form of: ± ( [ % of Reading ] + [ Counts ] )
Maximum voltage between any
terminal and earth ground:
600 V
Surge Protection: 6 kV peak per IEC 61010-1-95
W Fuse for A input: 11 A, 1000 V FAST Fuse
Display: Digital:
Bar Graph:
Frequency:
Capacitance:
6,000 counts, updates 4/sec
33 segments, updates 40/sec
9,999 counts
9,999 counts
Temperature: Operating:
Storage:
−10 °C to +50 °C
−30 °C to +60 °C
Temperature Coefficient: 0.1 x (specified accuracy)/ °C for <18 °C or > 28 °C
Electromagnetic Compatibility: Performance ≥ 3 V/m is not specified.
Relative Humidity: Noncondensing < 10 °C
0 % to 95 % @ 10 °C to 30 °C
0 % to 75 % @ 30 °C to 40 °C
0 % to 45 % @ 40 °C to 50 °C
Battery Life: Alkaline: 300 hours typical, without backlight
Size, with Holster (H x W x L): 4.6 cm x 9.6 cm x 16.0 cm
Weight: 350 g
Safety Compliance: ANSI/ISA-S82.01-1994, CSA C22.2 No 1010.1-92 and IEC 61010-1-95
Overvoltage Category III (CAT III), 600 V
Certifications: UL (3111), P, CSA, TÜV,
; (N10140)
Accuracy ± ( [ % of Reading ] + [ Counts ] )
Function Range Resolution Model 110 Model 111 Model 112
AC Volts
1,2
- True RMS
(50 Hz to 500 Hz)
6000 mV
3
6.000 V
60.00 V
600.0 V
1 mV
0.001 V
0.01 V
0.1 V
1.0 % + 3 1.0 % + 3 1.0 % + 3
DC Volts 6000 mV
3
6.000 V
60.00 V
600.0 V
1 mV
0.001 V
0.01 V
0.1 V
0.7 % + 2 0.7 % + 2 0.7 % + 2
Continuity 600 Ω 1 Ω Beeper guaranteed on < 20 Ω, guaranteed off > 250 Ω;
detects opens or shorts of 250 µs or longer.
Ohms 600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
40.00 MΩ
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
1. AC voltage ranges are specified from 5% of range to 100% of range.
2. Crest factor of ≤ 3 at full scale up to 300 V, decreasing linearly to crest factor ≤ 1.5 at 600 V.
3. The 6000 mV range can only be entered in Manual Range mode. Use the 6000 mVDC range with accessories.