1587 FC/1587/1577
Users Manual
8
Table 4. Rotary Switch Selections (cont.)
Switch
Position
Measurement Function
1587 FC
1587
1587T
1577
k
Temperature from - 40 °C to + 537 °C (- 40 °F to + 998 °F).
Celsius is the default temperature measurement unit. The temperature measurement you select is
retained in memory when the Meter is turned off.
o
Ohms from 0.1 Ω to 50 MΩ.
Capacitance from 1 nF to 9999 μF.
X
Continuity test. Beeper turns on at <25 Ω and turns off at >100 Ω.
O
Diode test. There is no ranging in this function. Displays  above 6.600 V.
c
AC mA from 3.00 mA to 400 mA (600 mA overload for 2 minutes maximum).
DC mA from 0.01 mA to 400 mA (600 mA overload for 2 minutes maximum).
a
INSULATION
Ohms from 0.01 MΩ to 2 GΩ. The last selected output voltage setting is retained in memory when the
Meter is turned off.
Ohms from 0.01 MΩ to 600 MΩ. The last selected output voltage setting is retained in memory when the
Meter is turned off.
Performs insulation test with:
50 V, 100 V, 250 V, 500 V (default), and 1000 V source
500 V (default) and 1000 V source
50 V (default) and 100 V source
Press the blue button to activate smoothing during insulation testing.
ГК Атлант Инжинирингофициальный представитель в РФ и СНГ
+7 (495) 109-02-08 sales@bbrc.ru www.bbrc.ru

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1587 MDT and is the answer not in the manual?

Fluke 1587 MDT Specifications

General IconGeneral
BrandFluke
Model1587 MDT
CategoryMultimeter
LanguageEnglish

Related product manuals