EasyManua.ls Logo

Fluke 1587 MDT - General Product Specifications; Technical Specifications Summary

Fluke 1587 MDT
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Insulation Multimeter
General Specifications
33
General Specifications
Maximum Voltage Applied to
any Terminal and Common .................................... 1000 V
Fuse Protection for mA input .................................. 0.44 A, 1000 V, IR 10 kA
Batteries .................................................................. Four AA batteries (NEDA 15A or IEC LR6)
Battery Life .............................................................. Meter use 1000 hours; Insulation test use: Meter can perform at least 1000 insulation tests with fresh
alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5
seconds on and 25 seconds off.
Size ......................................................................... 5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L)
Weight ..................................................................... 550 g (1.2 lb)
Altitude
Operating ........................................................ 2000 m
Storage ........................................................... 12,000 m
Over-Range Capability ........................................... 110 % of range except for capacitance which is 100 %
Frequency Overload Protection .............................. 10
7
V Hz
Storage Temperature.............................................. -40 °C to 60 °C (-40 °F to 140 °F)
Operating Temperature .......................................... -20 °C to 55 °C (-4 °F to 131 °F)
Temperature Coefficient ......................................... 0.05 x (specified accuracy) per °C for temperatures <18 °C or >28 °C (<64 °F or >82 °F)
Relative Humidity .................................................... Noncondensing
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)
Enclosure Protection............................................... IEC 60529: IP40 (non-operating)
Safety
IEC 61010-1 ................................................... Pollution Degree 2
IEC 61010-2-033 ............................................ CAT IV 600 V / CAT III 1000 V
ГК Атлант Инжинирингофициальный представитель в РФ и СНГ
+7 (495) 109-02-08 sales@bbrc.ru www.bbrc.ru

Related product manuals