1587 FC/1587/1577
Users Manual
20
Temperature (all 1587 Models)
The Meter measures the temperature of a type-K
thermocouple (included). Choose between degrees
Celsius (°C) or degrees Fahrenheit (°F).
1587 FC:
Press to toggle between °C or °F.
1587/1587T:
Press r to toggle between °C or °F.
Caution
To prevent possible damage to the Meter or
other equipment, remember that while the
Meter is rated for -40 °C to 537 °C (-40 °F to
998.0 °F), the included K-type thermocouple is
rated for 260 °C (500 °F). For temperatures out
of that range, use a higher rated thermocouple.
Warning
To prevent risk of electric shock do not
connect thermocouple to electrically live
circuits.
80BK-A Type K
Thermocouple
Probe
Vent
or
Pipe
°C
°F
bav09f.eps
Figure 3. Measuring Temperature