Insulation Specifications
Measurement Range
Model 1587, 1587 FC ................................. 0.01 MΩ to 2 GΩ
Model 1577 ................................................. 0.1 MΩ to 600 MΩ
Model 1587T ............................................... 0.01 MΩ to 100 MΩ
Test Voltages
Model 1587, 1587 FC ................................. 50, 100, 250, 500, 1000 V
Model 1577 ................................................. 500, 1000 V
Model 1587T ............................................... 50, 100 V
Test Voltage Accuracy ........................................ +20 %, -0 %
Short-Circuit Test Current ................................... 1 mA nominal
Auto Discharge ................................................... Discharge time <0.5 second for C = 1 μF or less
Live Circuit Detection .......................................... Inhibit test if terminal voltage > 30 V prior to initialization of test
Maximum Capacitive Load ................................. Operable with up to 1 μF load
ГК Атлант Инжинирингофициальный представитель в РФ и СНГ
+7 (495) 109-02-08 sales@bbrc.ru www.bbrc.ru

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1587 MDT and is the answer not in the manual?

Fluke 1587 MDT Specifications

General IconGeneral
BrandFluke
Model1587 MDT
CategoryMultimeter
LanguageEnglish

Related product manuals