Electrical Installation Tester
Performance Tests
21
280V
RMS
MAX
280V
RMS
MAX
CA T I
1000V
CA T II
600V
RMS MAX
20V PK
30A
RMS
MAX
20V PK
LO
LO - SENSE
, HI ,
mA
V
1500V PK
MAX
50V PK
MAX
20V PK
20V PK
N
L
PE
L1 L2 L3HES
OUTPUT
Z
L
, Z
GND
,
RCD
HI
LO
HI
LO
METER
METER
5320A MU LTIFUNCTION ELECTRICAL
V
A
COM
INPUT
Fluke 5320A
PE
(L2/Green)
L
(L1/Red)
gad002.eps
Figure 4. Touch Pad Sense Test
Memory Mode Test (1653B and 1654B Only)
Use the following test to verify the proper function of the UUT Memory Mode:
1. Turn the rotary switch to V.
2. Connect the PE and L input jacks of the Tester to the 5320A V, HIΩ, mA~ Output
terminals as shown in Figure 4.
3. Set the 5320A for the Volts function and apply 25 V, 50 Hz to the Tester.
4. Press UUT to enter Memory Mode.
5. Use < and the to set the data identity.
6. Press to save the data.
To recall the saved data:
7. Press to re-enter Memory Mode.
8. Use < and to set the data identity.
9. Press twice to recall the data. The recalled data should match the saved value.