Calibration Manual
24
Continuity Function
To verify accuracy of the UUT continuity function:
1. Turn the UUT rotary switch to CONTINUITY R
LO
.
2. Connect a shorting bar directly to the UUT PE and L input jacks.
3. Press and hold the UUT until the Tester displays a reading and the ZERO ∅
annunciator.
4. Compare the UUT display reading with the Display Reading Limits for step 1 of
Table 4.
5. Connect the measurement test leads to the UUT L and PE input jacks in a 4-wire
configuration as shown in Figure 6.
6. Connect the UUT test leads to the 5320A LOΩ Output terminals as shown in Figure
6.
7. Apply the 5320A Output values for the remaining steps in Table 4, and press the
UUT .
8. Compare each UUT display reading with the Display Limits in Table 4.