Electrical Installation Tester
Equipment Requirements
15
Equipment Requirements
The Tester’s recommended calibration cycle is one year. Table 2 lists the calibration
equipment required for the Tester. If the recommended models are unavailable, you can
substitute equipment with equivalent specifications.
Warning
For safe operation and maintenance of the product, have an
approved technician repair the product.
Table 2. Equipment Requirements
Equipment Description
Recommended
Model
Multifunction Electrical Tester
Calibrator
AC Voltage Range: 0 to 500 V ac
Accuracy: ±0.215 %
Frequency Range: 50 Hz
Insulation Resistance: 10 k to 900 M
Accuracy:
10 k to 200 M = ±0.413 %
200 M to 900 M = 2.5 %
Low Resistance: 2 to 1800
Accuracy: ±0.416 %
Loop Resistance: 100 m to 1.8 k
Accuracy:
100 m = ±5 %
1.8 = ±1 %
18 to 180 = ±0.5 %
500 to 1800 = ±1.5 %
RCD
Trip Current Measurement
Range: 10 to 1000 mA
Accuracy:
0.5 x 1 and 1 x 1 mode = ±2.5 %
1.4 x 1 and 2 x 1 mode = ±2.5 %
5 x 1 mode = ±2.5%
Trip time Measurement
Range: 30.0 to 500.0 ms
Accuracy: ±1.2 %
Fluke 5320A Calibrator
Decade Resistor Box Resistance: 18 , 180 , 1800
Accuracy: 0.75 % or better
Power: 0.5 W minimum
GenRad 1433
PC PC with Windows 98 or newer ---
IR adapter Cable --- Fluke P/N 1590638
IR Cable adapter holder FLK-1653-2014-OSP, 2043365, IR Adapter Fluke P/N 3270172
Zero Adapter Fluke-165XB-8001, Zero Adapter Fluke P/N 3301338