EasyManua.ls Logo

Fluke 165X - Insulation Resistance Measurement (R ISO )

Fluke 165X
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
165X
Users Manual
48
Live Circuit Detection Inhibits test if terminal voltage > 10 V ac detected prior to
initiation of test.
Insulation Resistance Measurement (R
ISO
)
Test Voltages
Model 1651 Model 1652 Model 1653
Accuracy of Test Voltage
(at rated test current)
500-1000 V 250-500-1000 V 50-100-250-500-1000 V +10%, -0%
Test Voltage
Insulation
Resistance Range Resolution Test Current Accuracy
50 V 10 k to 50 M 0,01 M 1 mA @ 50 k + (3%+3 digits)
100 k to 20 M 0,01 M 1 mA @ 100 k + (3%+3 digits) 100 V
20 M to 100 M 0,1 M 1 mA @ 100 k + (3%+3 digits)
250 V 100 k to 200 M 0,1 M 1 mA @ 250 k + (1,5%+3 digits)
100 k to 200 M 0,1 M 1 mA @ 500 k + (1,5%+3 digits) 500 V
200 M to 500 M 1 M 1 mA @ 500 k + 10%
100 k to 200 M 0,1 M 1 mA @ 1 M + (1,5%+3 digits) 1000 V
200 M to 1000 M 1 M 1 mA @ 1 M + 10%
The number of possible insulation tests with a fresh set of batteries is 2000.