Calibration Adjustment
5.7 Save Calibration Data And Exit 5
5-13
Proceed as follows:
If you did the Warming Up and Pre-Calibration successfully (section 5.5), and you
want to store the Pre-Calibration data before continuing with the Final Calibration:
Press
F5
YES.
When turning the Analyzer off and on again, it will show the message:
The instrument needs calibration.
Please contact your service center.
The calibration date and number will not be updated. You must continue with the Final
Calibration!
To return to the Maintenance mode, if you want to repeat the complete calibration:
Press
F4
NO.
Now press
F1
until the display shows Warming Up (CL 0200):IDLE, and
calibrate the Analyzer, starting at section 5.5.
If you want to exit and maintain the old calibration data:
Turn the Analyzer off.